The Tracheoesophageal Puncture Dilator is a tapered, solid silicone stent that prevents the tracheoesophageal passage from closing upon removal of a voice prosthesis. Available in two sizes.
Size | Product Code |
18 Fr for 16 Fr prosthesis | BE6050 |
22 Fr for 20 Fr prosthesis | BE2050 |
DEVICE CLEANING: Puncture Dilator The Puncture Dilator may be cleaned and reused by a single patient as follows: Clean the Puncture Dilator thoroughly with a mild detergent and rinse under a strong flush of tap water. Dry the dilator with sterile gauze. Once a dilator has been cleaned and dried completely, it should be stored in a clean, reclosable plastic bag until its next use. Always handle with clean hands and avoid touching portions of the dilator that will enter the puncture. Precaution: Never clean the device while the dilator is in position in the neck. Always clean the dilator after it has been removed.
INSTITUTION STERILISATION AND REUSE INSTRUCTIONS The devices of the Blom-Singer Dilation/Sizing System may be cleaned and steam sterilised for reuse by multiple patients as follows: Clean thoroughly with a mild detergent and rinse with water. Sterilize unwrapped in a standard autoclave at 132— 135°C for 3 minutes. Do not sterilize the device in the original packaging. The device may be reused to a limit of 100 times. Inspect the device routinely for structural damage. If there are tears, cracks, or structural damage, do not continue to use. Do not sterilise the sizer inserter
Brand |
---|
The information contained within this website is designed and intended for healthcare professionals only.
Endotherapeutics will not be liable for any actions taken in reliance of the information contained within the website. The information contained within the website does not constitute medical advice.
If you are a patient who requires treatment or management of a medical condition, please click No and you will be redirected to Endo Personal Care.